Z-2696-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 29, 2020
- Initiation Date
- June 30, 2020
- Termination Date
- January 25, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 28
Product Description
IntelliVue MX800 patient monitor, model no. 865240
Reason for Recall
Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.
Distribution Pattern
Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.
Code Information
Serial numbers DE50887321 DE50887434 DE50887128 DE50887025 DE50887196 DE50887429 DE50887125 DE50887043 DE50887303 DE50887420 DE50887106 DE50887164 DE50887322 DE50887607 DE50887061 DE50887179 DE50887425 DE50887427 DE50887142 DE50887446 DE50887152 DE50887064 DE50887445 DE50887137 DE50887063