Z-2710-2016 Class II Terminated
FDA device recall Z-2710-2016 was initiated by Becton Dickinson & Company on Aug 8, 2016 and is designated Class II. Reason for recall: Becton Dickinson is recalling various lot numbers of BD Vacutainer Push Button Blood Collection Set because of reports that specific lots exhibit separation of fron and rear barrels upon activation of… The recall status is terminated (terminated Nov 20, 2017). Affected quantity: 1,788,150 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 7, 2016
- Initiation Date
- Aug 8, 2016
- Termination Date
- Nov 20, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,788,150 units
Product description
BD Vacutainer Push Button Blood Collection Set, Sterile, Rx only, 0.6 x 19 mm x 305 mm, 23G x 3/4" x 12", REF 367342.
Reason for recall
Becton Dickinson is recalling various lot numbers of BD Vacutainer Push Button Blood Collection Set because of reports that specific lots exhibit separation of fron and rear barrels upon activation of the safety feature, which retracts the needle.
Distribution pattern
US and International
Code information
Lot number: 5355531, 5355533, 5356681, 5356682, 6006629, 6006630, 6006631, 6006633, 6006634 and 6006650.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2710-2016
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026