Z-2717-2026 Class II Ongoing

Recalled by CMR SURGICAL LIMITED — Cambridge, N/A

FDA device recall Z-2717-2026 was initiated by CMR SURGICAL LIMITED on June 9, 2026 and is designated Class II. Reason for recall: Secure Boot was mistakenly not enabled at manufacturing time, which presents a potential cybersecurity risk. The recall status is ongoing. Affected quantity: 2.

Recall Details

Product Type
Devices
Report Date
July 15, 2026
Initiation Date
June 9, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product Description

Brand Name: Versius Surgical System US Product Name: Vesius Surgeon Console US Model/Catalog Number: V-SS-1020 Software Version: Pratchett 1 version 62.1 Product Description: The Versius Surgical System is a modular, open console, software controlled, electromechanical system designed for surgeons to perform minimally invasive surgery.

Reason for Recall

Secure Boot was mistakenly not enabled at manufacturing time, which presents a potential cybersecurity risk.

Distribution Pattern

US Nationwide distribution in the state of California.

Code Information

UDI-DI 05060548211226 Model No V-SS-1020 Serial Numbers 66495, 66496 Software Revision Version 62.1 There is no expiry date for this device.