Z-2724-2024 Class II
FDA device recall Z-2724-2024 was initiated by Stryker Corporation on Jul 24, 2024 and is designated Class II. Reason for recall: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor… Affected quantity: 1226 units.
Recall details
- Product Type
- Devices
- Report Date
- Sep 4, 2024
- Initiation Date
- Jul 24, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1226 units
Product description
Stryker iBur 2.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-012-020D
Reason for recall
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.
Code information
GTIN: 07613327501261 Lot Numbers: 22237017 22241017 22334017 23085017 23179017 23213017 24003017 24015017