Z-2746-2016 Class II Terminated
FDA device recall Z-2746-2016 was initiated by Maquet Cardiovascular Us Sales, Llc on Aug 12, 2016 and is designated Class II. Reason for recall: Post Market Surveillance cases reported to Maquet showing instances in which the welded joint between the square mount and the first fixable swivel joint of Maquet's Connecting Bracket (Fixture)… The recall status is terminated (terminated May 4, 2017). Affected quantity: 19 units (US) and 1975 units (OUS).
Recall details
- Product Type
- Devices
- Report Date
- Sep 14, 2016
- Initiation Date
- Aug 12, 2016
- Termination Date
- May 4, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 19 units (US) and 1975 units (OUS)
Product description
The Connecting Bracket (Fixture) Device (1002.65A0) is designed as a connecting element to support ancillary head rests on a Maquet operating room table immediately before, during and after surgical interventions as well as for examination and treatment. The fixture is used to adapt a range of different head and horseshoe head rests with a central square mount
Reason for recall
Post Market Surveillance cases reported to Maquet showing instances in which the welded joint between the square mount and the first fixable swivel joint of Maquet's Connecting Bracket (Fixture) Device is broken.
Distribution pattern
Nationwide Distribution to CA, FL, IA, IN, MO, UT and WI
Code information
Part Number/Catalog Number 1002.64A0
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2746-2016
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab