Z-2747-2017 Class II Terminated
FDA device recall Z-2747-2017 was initiated by Becton Dickinson & Company on Sep 16, 2016 and is designated Class II. Reason for recall: BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required. The recall status is terminated (terminated Mar 20, 2018). Affected quantity: 22,400 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 26, 2017
- Initiation Date
- Sep 16, 2016
- Termination Date
- Mar 20, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 22,400 units
Product description
BD 5ml Syringe Luer-Lok" Tip with Blunt Fill Needle 18G x 1 1/2 (1.2mm x 40mm) Catalog #305062
Reason for recall
BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.
Distribution pattern
Nationwide Distribution
Code information
6204878 2021-07