Z-2748-2017 Class II Terminated
FDA device recall Z-2748-2017 was initiated by Becton Dickinson & Company on Sep 16, 2016 and is designated Class II. Reason for recall: BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required. The recall status is terminated (terminated Mar 20, 2018). Affected quantity: 29,500 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 26, 2017
- Initiation Date
- Sep 16, 2016
- Termination Date
- Mar 20, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 29,500 units
Product description
BD SafetyGlide" Injection Needle Catalog #305917 The SafetyGlide " needle is intended to be used for general purpose injection, aspiration of fluids from parts of the body below the surface of the skin, injection into intravenous ports and tubings external to the patient. The needlestick prevention feature of the device contains a mechanism that covers the needlepoint after use. In the activated position the needle cover guards against accidental needle sticks during normal handling and disposal of the used needle/syringe combination
Reason for recall
BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.
Distribution pattern
Nationwide Distribution
Code information
6022838 2021-01