Z-2753-2017 Class II Terminated
FDA device recall Z-2753-2017 was initiated by Becton Dickinson & Company on Sep 16, 2016 and is designated Class II. Reason for recall: BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required. The recall status is terminated (terminated Mar 20, 2018). Affected quantity: 1,528,000 units.
Recall details
- Product Type
- Devices
- Report Date
- Jul 26, 2017
- Initiation Date
- Sep 16, 2016
- Termination Date
- Mar 20, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,528,000 units
Product description
BD 3ml Syringer Luer-Lok Tip Catalog #309604 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
Reason for recall
BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.
Distribution pattern
Nationwide Distribution
Code information
6061746 2021-02 6082982 2021-02 6089821 2021-03 6117858 2021-04 6117863 2021-04 6117889 2021-04 6117891 2021-04 6146962 2021-04 6173697 2021-05 6173767 2021-06 6173779 2021-06 6208614 2021-07