Z-2756-2016 Class II Terminated
FDA device recall Z-2756-2016 was initiated by Sorin Group USA, Inc. on Aug 12, 2016 and is designated Class II. Reason for recall: Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90… The recall status is terminated (terminated Mar 27, 2017). Affected quantity: 200 units worldwide.
Recall details
- Product Type
- Devices
- Report Date
- Sep 14, 2016
- Initiation Date
- Aug 12, 2016
- Termination Date
- Mar 27, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 200 units worldwide
Product description
Aortic Arch Cannula and Coronary Artery Perfusion Cannula with Ballon (part number NA-55X6). The Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect. The Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip
Reason for recall
Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.
Distribution pattern
Worldwide Distribution - US including IL, MO, TX. Internationally to Canada and Italy.
Code information
Lot Numbers: 1526500032, 1529300026, 1602600042.