Z-2757-2016 Class II Terminated

Recalled by Sorin Group USA, Inc. — Arvada, CO

FDA device recall Z-2757-2016 was initiated by Sorin Group USA, Inc. on Aug 12, 2016 and is designated Class II. Reason for recall: The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement. The recall status is terminated (terminated Mar 27, 2017). Affected quantity: 1205 units worldwide.

Recall details

Product Type
Devices
Report Date
Sep 14, 2016
Initiation Date
Aug 12, 2016
Termination Date
Mar 27, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1205 units worldwide

Product description

Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided for with a vinyl female luer connector bonded proximally and a vinyl cuff bonded distally. The cuff is pre-molded to its inflated configuration and is purposely positioned directly over a hole generated in the extruded tube for purposes of receiving the infused cardioplegia solution. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.

Reason for recall

The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.

Distribution pattern

Worldwide Distribution - US including IL, MO, TX. Internationally to Canada and Italy.

Code information

Lot Numbers: 1425900137, 1433900079, 1433900080, 1435300088.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2757-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2757-2016. Report a data issue: hello@regdatalab.com.