Z-2764-2019 Class III Terminated
FDA device recall Z-2764-2019 was initiated by ELITech Group B.V. on Jul 12, 2016 and is designated Class III. Reason for recall: Software malfunction; The action being taken due to the device becoming inoperable due to software lockup. The recall status is terminated (terminated Feb 4, 2022). Affected quantity: 89 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 9, 2019
- Initiation Date
- Jul 12, 2016
- Termination Date
- Feb 4, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 89 units
Product description
Viva-ProE Systems, Model # 6003-670, UDI: 03661540600357 - Product Usage: The ELITech Clinical Systems Selectra ProM is an automated clinical chemistry system intended for use in clinical laboratories. It is intended to be used for a variety of assay methods that have been applied to spectrophotometric and electrochemical techniques. The system has two core modules: one consisting of a spectrophoto-metric system for measurement of analytes using spectrophotometric techniques, such as end point, rate and turbidometric assays. The second module is an electro-meter used for measurement of electrolytes.
Reason for recall
Software malfunction; The action being taken due to the device becoming inoperable due to software lockup.
Distribution pattern
US Nationwide distribution including the state of NY.
Code information
1. 14-1301 2. 14-1304 3. 14-1308 4. 14-1314 5. 14-1315 6. 14-1323 7. 14-1324 8. 14-1336 9. 15-1328 10. 15-1332 11. 15-1333 12. 15-1334 13. 15-1335 14. 15-1336 15. 15-1337 16. 15-1338 17. 15-1339 18. 16-1301 19. 16-1302 20. 16-1304 21. 16-1305 22. 16-1312 23. 17-1335
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2764-2019
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
Need the full FDA record for ELITech Group B.V.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.