Z-2766-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 9, 2019
- Initiation Date
- June 12, 2017
- Termination Date
- October 30, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- N/A
Product Description
Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
Reason for Recall
Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
Distribution Pattern
Worldwide distribution - US Nationwide distribution.
Code Information
Serial numbers less than xxx25525056 Updated serial numbers on 10/12/2019: Performa II: Between XXX20000000 and XXX25525056 Updated 10/18/2019: Performa II: from XXX20000000 to XXX25525056