Z-2769-2015 Class II Terminated

Recalled by Applied Medical Resources Corp — Rancho Santa Margarita, CA

FDA device recall Z-2769-2015 was initiated by Applied Medical Resources Corp on Jul 31, 2015 and is designated Class II. Reason for recall: Potential to improperly characterize tissue within the jaws, which may result in nonoptimal power delivery. The recall status is terminated (terminated Nov 20, 2015). Affected quantity: 101 units (U.S) and 38 units (International).

Recall details

Product Type
Devices
Report Date
Sep 23, 2015
Initiation Date
Jul 31, 2015
Termination Date
Nov 20, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
101 units (U.S) and 38 units (International)

Product description

Voyant Open Fusion Device Models: EB040 and EB040+ Product Usage: The Applied Voyant Open Fusion Device is designed for use with the Voyant Electrosurgical Generator. The device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 5mm (EB040+) or 7mm (EB040) in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The system delivers the energy required to fuse the tissue between the jaws of the device by monitoring voltage and current throughout the seal cycle. The device also features a useractuated blade for the transection of sealed tissue.

Reason for recall

Potential to improperly characterize tissue within the jaws, which may result in nonoptimal power delivery.

Distribution pattern

Worldwide Distribution - U.S Nationwide in the states of:FL, GA, NV. and countries of: France, Netherlands, Spain, and United Kingdom.

Code information

Model Number: EB040 Lot Numbers: 1243964, 1243975, 1244125 Model Number: EB040+ Lot Numbers: 1245512, 1245513, 1246241, 1246508, 1247079, 1247566, 1248102, 1249264

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2769-2015
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Applied Medical Resources Corp? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Applied Medical Resources Corp: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Applied Medical Resources Corp →

Source: openFDA record for Z-2769-2015. Report a data issue: hello@regdatalab.com.