Z-2797-2017 Class II Terminated

Recalled by Sorin Group USA, Inc. — Arvada, CO

FDA device recall Z-2797-2017 was initiated by Sorin Group USA, Inc. on Jun 29, 2017 and is designated Class II. Reason for recall: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a… The recall status is terminated (terminated Apr 12, 2021). Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Aug 9, 2017
Initiation Date
Jun 29, 2017
Termination Date
Apr 12, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

NON-STERILE SAMPLE, SORIN GROUP, SMART CARDIOPLEGIA, CARDIOPLEGIA PACK, REF 627366901, 1 EA, NOT FOR CLINICAL USE

Reason for recall

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Distribution pattern

Nationwide

Code information

Lot Number 1623100001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2797-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2797-2017. Report a data issue: hello@regdatalab.com.