Z-2799-2020 Class II Terminated

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai, N/A

Recall Details

Product Type
Devices
Report Date
August 19, 2020
Initiation Date
July 16, 2020
Termination Date
July 7, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit distributed to Texas

Product Description

uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Reason for Recall

1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions.

Distribution Pattern

US Nationwide distribution including in the state of Texas.

Code Information

Serial No. F00002; GTIN 06971576832040