Z-2799-2020 Class II Terminated

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai, N/A

FDA device recall Z-2799-2020 was initiated by Shanghai United Imaging Healthcare Co., Ltd. on Jul 16, 2020 and is designated Class II. Reason for recall: 1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external… The recall status is terminated (terminated Jul 7, 2022). Affected quantity: 1 unit distributed to Texas.

Recall details

Product Type
Devices
Report Date
Aug 19, 2020
Initiation Date
Jul 16, 2020
Termination Date
Jul 7, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit distributed to Texas

Product description

uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Reason for recall

1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions.

Distribution pattern

US Nationwide distribution including in the state of Texas.

Code information

Serial No. F00002; GTIN 06971576832040

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2799-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Shanghai United Imaging Healthcare Co., Ltd.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Shanghai United Imaging Healthcare Co., Ltd.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Shanghai United Imaging Healthcare Co… →

Source: openFDA record for Z-2799-2020. Report a data issue: hello@regdatalab.com.