Z-2812-2020 Class II Terminated

Recalled by Philips North America, LLC — Andover, MA

FDA device recall Z-2812-2020 was initiated by Philips North America, LLC on Jun 18, 2020 and is designated Class II. Reason for recall: A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED… The recall status is terminated (terminated Jan 10, 2023). Affected quantity: 9,705 units.

Recall details

Product Type
Devices
Report Date
Aug 26, 2020
Initiation Date
Jun 18, 2020
Termination Date
Jan 10, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9,705 units

Product description

Phillips Module Charger, Product #: 989803191031 - Product Usage: This Module Charger is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.

Reason for recall

A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.

Distribution pattern

World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed.

Code information

Product #: 989803191031 Manufacturing date codes: prior to 2020-03

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2812-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America, LLC →

Source: openFDA record for Z-2812-2020. Report a data issue: hello@regdatalab.com.