Z-2819-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 26, 2020
- Initiation Date
- March 26, 2020
- Termination Date
- August 2, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3
Product Description
Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Reason for Recall
In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.
Distribution Pattern
US Nationwide distribution including in the state of TX.
Code Information
UDI: 06971576832026, Software Version: R001.3.0.0.750505, Serial Numbers: 200017, 200023, 200024