Z-2819-2020 Class II Terminated

Recalled by Shanghai United Imaging Healthcare Co., Ltd. — Shanghai, N/A

Recall Details

Product Type
Devices
Report Date
August 26, 2020
Initiation Date
March 26, 2020
Termination Date
August 2, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product Description

Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Reason for Recall

In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.

Distribution Pattern

US Nationwide distribution including in the state of TX.

Code Information

UDI: 06971576832026, Software Version: R001.3.0.0.750505, Serial Numbers: 200017, 200023, 200024