Z-2822-2015 Class II Terminated

Recalled by MicroAire Surgical Instruments, LLC — Charlottesville, VA

FDA device recall Z-2822-2015 was initiated by MicroAire Surgical Instruments, LLC on Jun 12, 2015 and is designated Class II. Reason for recall: There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails. The recall status is terminated (terminated May 23, 2016). Affected quantity: 168, ASP-1021 LipoTower with Aspiration Pump and Tumescent Pump.

Recall details

Product Type
Devices
Report Date
Oct 7, 2015
Initiation Date
Jun 12, 2015
Termination Date
May 23, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
168, ASP-1021 LipoTower with Aspiration Pump and Tumescent Pump

Product description

ASP-1021 LipoTower with Aspiration Pump and Tumescent Pump. Medical use.

Reason for recall

There is potential risk of electrical shock if lead wires inside the device become detached and the ground connection fails.

Distribution pattern

Nationwide Distribution.

Code information

Serial numbers 14101241, 4091238, 15011258, 15051279, 13101158, 15051275, 14021190, 12121111, 15051274, 14071221, 13081149, 13031123, 14021187, 14121254, 14081231, 13121173, 13111167, 13101164, 13121172, 14091234, 14111248, 14031196, 13061142, 15031264, 13121171, 13021116, 15051276, 13041131, 13101160, 14051210, 15041271, 14071222, 14051208, 15021260, 13111166, 13101163, 14111246, 13111168, 13101158, 14061212, 13081152, 13121177, 13061145, 15051280, 15061283, 14041200, 13021114, 14011184, 14031193, 13111165, 14011182, 14111249, 14101242, 14011181, 15041272, 13101156, 13041126, 14101240, 14021188, 12111108, 13061141, 12121112, 13061140, 13041128, 13081150, 14031195, 15051281, 14031197, 14081228, 14101243, 14031199, 14081232, 14071226, 13081148, 14111244, 14091238, 15051279, 13091154, 13121179, 14121252, 15041268, 14071225, 14071223, 15031266, 13031122, 13061144, 14021186, 14091233, 13101161, 14031194, 15041273, 13121170, 13021120, 13011113, 13121174, 14041201, 15021261, 14011183, 13061143, 13121175, 15021262, 15011259, 13101157, 14071227, 13091155, 13111169, 14121251, 14061211, 14071216, 14041207, 13041124, 14021189, 14011185, 14031198, 12101105, 14071224, 14041203, 14071220, 13051132, 14091239, 15041269, 14051209, 15021263, 15011257, 13081147, 13031121, 14101241, 15011258, 13041125, 13081151, 13101159, 13121176, 12121109, 14061213, 13121178, , 13101162, 14111245, 13061138, 14081229, 14021192, 14091237, 14121253, 13061139, 14121255, 15051278, 15051277, 14071217, 15011256, 13021118, 13071146, 13051135, 13051136, 13051137, 14071219, 14081230, 13041127, 12101106, 14061215, 12101107, 14061214, 15041267, 14041206, 14041202, 13041130, 13041129, 14021191, 13051133, 13051134, 14011180, 14091235, 14091236, 14111247, and 14111250.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2822-2015
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for MicroAire Surgical Instruments, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works