Z-2845-2018 Class II Terminated

Recalled by Philips Medical Systems Gmbh, DMC — Hamburg

Recall Details

Product Type
Devices
Report Date
August 29, 2018
Initiation Date
June 27, 2018
Termination Date
April 30, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity

Product Description

DigitalDiagnost Release 3.1 (Stitching Patient Support) 712026

Reason for Recall

The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.

Distribution Pattern

US Nationwide; International to 66 countries

Code Information

Serial Numbers: SN13000495 SN14000058 SN13000463 SN14000051 SN14000214 SN14000047 SN14000077 SN13000484 SN14000082 SN14000163 SN14000133 SN14000134 SN14000134 SN14000023 SN14000041