Z-2862-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 29, 2018
- Initiation Date
- June 27, 2018
- Termination Date
- April 30, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
Product Description
DigitalDiagnost Release 3.1 (Stitching Patient Support) 712026
Reason for Recall
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Distribution Pattern
US Nationwide; International to 66 countries
Code Information
Serial Numbers: SN14000256 SN14000250 SN15000024 SN14000170 SN14000052 SN14000224 SN15000039 SN14000144 SN14000253 SN14000239 SN14000186 SN14000075 SN14000193 SN14000254 SN15000001 SN14000089 SN14000223 SN14000267 SN14000238 SN14000164 SN14000165 SN14000200 SN14000005 SN14000040 SN15000015 SN15000029 SN14000120