Z-2850-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 29, 2018
- Initiation Date
- June 27, 2018
- Termination Date
- April 30, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
Product Description
DigitalDiagnost 4 Flex / Value (Stitching Patient Support) 712028
Reason for Recall
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Distribution Pattern
US Nationwide; International to 66 countries
Code Information
Serial Numbers: SN17000283 SN15000078 SN15000115 SN17000326 SN15000114 SN17000171 SN17000058 SN17000190 SN15000016 SN15000316 SN17000027 SN14000047 SN17000024 SN15000015 SN17000342 SN14000108