Z-2850-2018 Class II Terminated

Recalled by Philips Medical Systems Gmbh, DMC — Hamburg

FDA device recall Z-2850-2018 was initiated by Philips Medical Systems Gmbh, DMC on Jun 27, 2018 and is designated Class II. Reason for recall: The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient. The recall status is terminated (terminated Apr 30, 2024).

Recall Details

Product Type
Devices
Report Date
Aug 29, 2018
Initiation Date
Jun 27, 2018
Termination Date
Apr 30, 2024
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

DigitalDiagnost 4 Flex / Value (Stitching Patient Support) 712028

Reason for Recall

The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.

Distribution Pattern

US Nationwide; International to 66 countries

Code Information

Serial Numbers: SN17000283 SN15000078 SN15000115 SN17000326 SN15000114 SN17000171 SN17000058 SN17000190 SN15000016 SN15000316 SN17000027 SN14000047 SN17000024 SN15000015 SN17000342 SN14000108