Z-2907-2016 Class II Terminated
FDA device recall Z-2907-2016 was initiated by Integra LifeSciences Corp. on Aug 23, 2016 and is designated Class II. Reason for recall: During final sampling inspection, one of the lots was discovered with a total incomplete packaging seal of the sterile product. The recall status is terminated (terminated Mar 29, 2017). Affected quantity: 18.
Recall details
- Product Type
- Devices
- Report Date
- Oct 12, 2016
- Initiation Date
- Aug 23, 2016
- Termination Date
- Mar 29, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 18
Product description
Integral Lumbar Drainage Set, REF 910420, Sterile EO, Rx only, primary packaged in a PVC blister sealed with Tyvek sheet and secondary packaged in to Tyvek pouch. Internal drainage set is primary packaged in a Tyvek pouch and secondary packaged in to Tyvek pouch, One (1) sealed package of lumbar catheter and Internal drainage set are finally packed in a cardboard box. Intended for CSF external drainage
Reason for recall
During final sampling inspection, one of the lots was discovered with a total incomplete packaging seal of the sterile product.
Distribution pattern
Nationwide Distribution to KS only
Code information
Catalog Number 910420, Lot #0195878 with expiry 10/2018
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2907-2016
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026