Z-2908-2020 Class II Terminated

Recalled by Neurovision Medical Products Inc — Ventura, CA

FDA device recall Z-2908-2020 was initiated by Neurovision Medical Products Inc on Jul 21, 2020 and is designated Class II. Reason for recall: The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could… The recall status is terminated (terminated Apr 30, 2021). Affected quantity: 1,058.

Recall details

Product Type
Devices
Report Date
Sep 2, 2020
Initiation Date
Jul 21, 2020
Termination Date
Apr 30, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,058

Product description

Cobra x5 2-Ch EMG ET Tube, 8.0mm, REF: LTE700DCL-5, Rx Only, Sterile EO UDI: B006LTE700DCL52 - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.

Reason for recall

The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.

Distribution pattern

Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.

Code information

Kit Lots: 022019A 052919B 053119A 080819A 022519D 081319E 082619A 082719B 021220E 022820C 060120A 110819A 111919F 030519A 030819A 091019B

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2908-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Neurovision Medical Products Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Neurovision Medical Products Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Neurovision Medical Products Inc →

Source: openFDA record for Z-2908-2020. Report a data issue: hello@regdatalab.com.