Z-2909-2016 Class II Terminated
FDA device recall Z-2909-2016 was initiated by Integra LifeSciences Corp. on Aug 23, 2016 and is designated Class II. Reason for recall: During final sampling inspection, one of the lots was discovered with a total incomplete packaging seal of the sterile product. The recall status is terminated (terminated Mar 29, 2017). Affected quantity: 14.
Recall details
- Product Type
- Devices
- Report Date
- Oct 12, 2016
- Initiation Date
- Aug 23, 2016
- Termination Date
- Mar 29, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14
Product description
Integra Ventricular Catheter Accessory Kit, REF 951303, Sterile EO, Rx only, packaged in a PVC tray, then secondary package in to Tyvek pouch and tertiary packaged in to Tyvek pouch. Sealed pouches are packaged in a cardboard box. F8 35cm Silicone Elastomer Ventricular Catheter for use with CSF external drainage systems
Reason for recall
During final sampling inspection, one of the lots was discovered with a total incomplete packaging seal of the sterile product.
Distribution pattern
Nationwide Distribution to KS only
Code information
Catalog Number 951303, Lot #0195786 with expiry 05/2019