Z-2910-2026 Class II Ongoing
FDA device recall Z-2910-2026 was initiated by Freudenberg Medical, LLC on Jun 17, 2026 and is designated Class II. Reason for recall: Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large… The recall status is ongoing. Affected quantity: 40 units.
Recall details
- Product Type
- Devices
- Report Date
- Aug 12, 2026
- Initiation Date
- Jun 17, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 40 units
Product description
Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF
Reason for recall
Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.
Distribution pattern
US Nationwide distribution in the states of IN, OH, AZ, WA, FL, MO, KY, IN, NY and the countries of GBR, FRA.
Code information
Lot #0000369205/UDI: 00813307014232
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2910-2026
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026