Z-2911-2026 Class II Ongoing

Recalled by Contego Medical, Inc — Raleigh, NC

FDA device recall Z-2911-2026 was initiated by Contego Medical, Inc on Jun 29, 2026 and is designated Class II. Reason for recall: Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath… The recall status is ongoing. Affected quantity: 5069.

Recall details

Product Type
Devices
Report Date
Aug 12, 2026
Initiation Date
Jun 29, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5069

Product description

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40

Reason for recall

Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.

Distribution pattern

US Nationwide distribution in the states of IL, FL, NC, WI, KY, AL, VA, DE, NJ, PA, IN, MI, CA.

Code information

REF/UDI-DI/Lot: NG-0630-140-2/00851616007583/Z2460578D,Z2460740B, Z2460872C, Z2460909F, Z2460948D, Z2460957Q; NG-0640-140-2/00851616007590/Z2460578E, Z2460647S, Z2460706A, Z2460746C, Z2460774A, Z2460865C, Z2460872B, Z2460901B, Z2460909D, Z2460909K, Z2460945H, Z2460948I, Z2460963N; NG-0730-140-2/00851616007606/Z2460578F, Z2460662A, Z2460740A , Z2460774F, Z2460850A, Z2460900B, Z2460945F, Z2460957R; NG-0740-140-2/00851616007613/Z2460578G, Z2460635C, Z2460656A, Z2460707D, Z2460733A, Z2460766A, Z2460774B, Z2460780A , Z2460865A, Z2460871A, Z2460872D, Z2460900D, Z2460909B, Z2460909I, Z2460948F, Z2460963T; NG-NV-6-30/00851616007644/Z2460776A, Z2460893A, Z2460909G, Z2460948E; NG-NV-6-40/00851616007651/Z2460776B, Z2460865D, Z2460865E, Z2460901C, Z2460909E, Z2460909J, Z2460957P, Z2460963K, Z2460963P; NG-NV-7-30/00851616007637/Z2460776C, Z2460789H, Z2460850B, Z2460900C, Z2460916A, Z2460948H, Z2460998A; NG-NV-7-40/00851616007620/Z2460776D, Z2460878A, Z2460872A, Z2460872E, Z2460901A, Z2460909C, Z2460909H, Z2460945G, Z2460948G, Z2460966B, Z2460963J;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2911-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.