Z-2921-2020 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-2921-2020 was initiated by Siemens Medical Solutions USA, Inc on Jul 24, 2020 and is designated Class II. Reason for recall: The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment… The recall status is terminated (terminated Nov 2, 2021). Affected quantity: 110 units US.

Recall details

Product Type
Devices
Report Date
Sep 9, 2020
Initiation Date
Jul 24, 2020
Termination Date
Nov 2, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
110 units US

Product description

Siemens Sensis Vibe Hemo system in combination with the MicroPodTM EtCO2 module-diagnostic and administrative tool supporting hemodynamic catheterization and/or electrophysiology studies, for cardiac as well as interventional Radiology.

Reason for recall

The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning and all vital signs are lost

Distribution pattern

US Nationwide distribution.

Code information

Serial Numbers: 100413 100414 100417 101016 101017 101019 101020 101022 101029 101031 101036 100263 100416 102008 101083 102138 102019 102021 102025 102029 102145 102147 100346 100349 102199 100095 100097 100103 100115 100116 100117 100120 100121 102158 102167 102168 102172 102272 102068 102069 102078 102080 102093 102221 100163 100165 100306 100309 101056 101099 101100 101101 101060 101068 101069 101070 101071 101072 100064 100422 100279 100408 100410 100411 100412 101043 102162 102285 102286 102183 100231 102060 102062 102049 102051 102057 102030 102031 101008 101093 101094 100175 100176 102226 101097 102269 102273 102276 102278 102279 102283 102193 102155 102157 102174 102182 102181 102214 102216 102219 102222 102237 102242 100391 101089 101095 102274 101018 102143 102091 9/01/20 Expanded S/N: 11007641 11007641 11007641 ***9/28/20 Expanded S/N*** 103053 103040 103042 103008

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2921-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA, Inc →

Source: openFDA record for Z-2921-2020. Report a data issue: hello@regdatalab.com.