Z-2947-2018 Class II Terminated
FDA device recall Z-2947-2018 was initiated by Civco Medical Instruments Co. Inc. on Jul 30, 2018 and is designated Class II. Reason for recall: The assembly of the needle guides were found to be either missing adhesive or did not have adequate adhesive to hold the assembled cannula in place on the needle guide. The recall status is terminated (terminated Sep 16, 2020). Affected quantity: 3 units.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 5, 2018
- Initiation Date
- Jul 30, 2018
- Termination Date
- Sep 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 units
Product Description
CIVCO Needle Guide, REF 676-150, QTY 24, STERILE EO
Reason for Recall
The assembly of the needle guides were found to be either missing adhesive or did not have adequate adhesive to hold the assembled cannula in place on the needle guide.
Distribution Pattern
United States, Austria, Belgium, Denmark, France, Germany, Poland, Spain, Switzerland and United Kingdom
Code Information
UDI 00841436107334, Lot numbers: A035054, A042207