Z-2951-2018 Class II Terminated
FDA device recall Z-2951-2018 was initiated by Haemonetics Corporation on Jul 13, 2018 and is designated Class II. Reason for recall: OrthoPAT¿ reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the… The recall status is terminated (terminated Jun 16, 2020). Affected quantity: 348 units.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 5, 2018
- Initiation Date
- Jul 13, 2018
- Termination Date
- Jun 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 348 units
Product Description
Haemonetics¿ OrthoPAT¿ System-OPAT QuickConnect Reservoirs. Product Code: OPT-R-1000 The OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures
Reason for Recall
OrthoPAT¿ reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
Distribution Pattern
US Nationwide Distribution
Code Information
Lot # 1117066
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2951-2018
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab