Z-2955-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 23, 2020
- Initiation Date
- August 4, 2020
- Termination Date
- August 12, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 units
Product Description
Integra Padgett Electric Dermatome Set - Product Usage: are intended for use in the cutting of skin grafts for use in plastic surgery and/or burn treatment. These instruments are intended only for professional use in hospitals and similar medical facilities, where the patient will be under the supervision of trained personnel.
Reason for Recall
lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of MN, TX, CA, UT, WV, FL, IN, GA, AZ, and the countries of Latin America ( CR, MX, TT, EC, PA ), Canada, Europe (FR, AT, BE, ES).
Code Information
Catalog Number: 3539900 Catalog Description: Electric Dermatome Set Model #: S6 Lot/Serial #: S-594-6 NOT SOLD IN USA