Z-2960-2018 Class II Terminated

Recalled by Bayer Medical Care, Inc. — Indianola, PA

FDA device recall Z-2960-2018 was initiated by Bayer Medical Care, Inc. on Jun 28, 2018 and is designated Class II. Reason for recall: The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may… The recall status is terminated (terminated Nov 26, 2018). Affected quantity: 2.

Recall details

Product Type
Devices
Report Date
Sep 12, 2018
Initiation Date
Jun 28, 2018
Termination Date
Nov 26, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product description

Overhead Counterpoise Systems (OCS), OCS CONF,STD CEIL MNT, STD HORZ ARM, Catalog # OCS 115A, Material Number 59880253

Reason for recall

The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.

Distribution pattern

The devices were distributed to the following US states: OH, TX, and WV.

Code information

Serial Numbers: 11051103902, 305746202

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2960-2018
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Bayer Medical Care, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2960-2018. Report a data issue: hello@regdatalab.com.