Z-2970-2018 Class I Terminated

Recalled by Pro-Med Instruments Gmbh — Freiburg Im Breisgau, N/A

FDA device recall Z-2970-2018 was initiated by Pro-Med Instruments Gmbh on Jul 5, 2018 and is designated Class I. Reason for recall: The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the… The recall status is terminated (terminated Nov 19, 2020). Affected quantity: 30 units.

Recall details

Product Type
Devices
Report Date
Sep 26, 2018
Initiation Date
Jul 5, 2018
Termination Date
Nov 19, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
30 units

Product description

DORO LUCENT Base Unit, REF 1101.021 Headrest System used in open and percutaneous craniotomies and spinal surgeries

Reason for recall

The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.

Distribution pattern

Worldwide distribution to US states of CA, TX, TN, MS, GA, WV, and OH, Qatar, India, UK, Singapore, Australia, Germany, South Africa, Italy, and Japan

Code information

Serial Numbers: 1704002 1704004 1704005 1704007 1704011 1704041 1704010 1704019 1704012 1704014 1704016 1704017 1704021 1704024 1704025 1704036 1704038 1704018 1704008 1704034 1704026 1704027 1704028 1704029 1704030 1704032 1704035 1704006 1704037 1704039

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2970-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Pro-Med Instruments Gmbh? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Pro-Med Instruments Gmbh: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Pro-Med Instruments Gmbh →

Source: openFDA record for Z-2970-2018. Report a data issue: hello@regdatalab.com.