Z-2976-2024 Class II Ongoing

Recalled by PACIFIC LASERTECH, LLC — San Marcos, CA

FDA device recall Z-2976-2024 was initiated by PACIFIC LASERTECH, LLC on Mar 1, 2024 and is designated Class II. Reason for recall: These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master… The recall status is ongoing. Affected quantity: 4.

Recall details

Product Type
Devices
Report Date
Sep 18, 2024
Initiation Date
Mar 1, 2024
Voluntary/Mandated
FDA Mandated
Product Quantity
4

Product description

25-LHP-928, HeNe Laser System, 25 mW

Reason for recall

These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.

Distribution pattern

US Nationwide Distribution

Code information

25-LHP-928

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2976-2024
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for PACIFIC LASERTECH, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2976-2024. Report a data issue: hello@regdatalab.com.