Z-2976-2024 Class II Ongoing

Recalled by PACIFIC LASERTECH, LLC — San Marcos, CA

Recall Details

Product Type
Devices
Report Date
September 18, 2024
Initiation Date
March 1, 2024
Termination Date
N/A
Voluntary/Mandated
FDA Mandated
Product Quantity
4

Product Description

25-LHP-928, HeNe Laser System, 25 mW

Reason for Recall

These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.

Distribution Pattern

US Nationwide Distribution

Code Information

25-LHP-928