Z-2977-2018 Class II Terminated

Recalled by Medline Industries, Inc. — Waukegan, IL

FDA device recall Z-2977-2018 was initiated by Medline Industries, Inc. on Apr 12, 2017 and is designated Class II. Reason for recall: Multiple surgical packs were assembled with a non-sterile CHG disk labeled as BIOPATCH" (kit component #52655) attached to the outside of the sterile pack. This may pose potential risk of infection… The recall status is terminated (terminated Aug 5, 2019). Affected quantity: 144 kits.

Recall details

Product Type
Devices
Report Date
Sep 12, 2018
Initiation Date
Apr 12, 2017
Termination Date
Aug 5, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
144 kits

Product description

PICC Tray, STERILE, 12 packs each per case.

Reason for recall

Multiple surgical packs were assembled with a non-sterile CHG disk labeled as BIOPATCH" (kit component #52655) attached to the outside of the sterile pack. This may pose potential risk of infection. However, the kits have been in distribution since March 2016. Many of these kits have been used with no reported infections or adverse patient events. In addition, the BIOPATCH is impregnated with CHG, a well-known antiseptic agent with broad-spectrum anti-microbial and antifungal activity.

Distribution pattern

CA, NM

Code information

Item number: DYNJ49177C. Lots: 16HB4700, 16UB5076, 17MB6837

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2977-2018
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medline Industries, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medline Industries, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medline Industries, Inc. →

Source: openFDA record for Z-2977-2018. Report a data issue: hello@regdatalab.com.