Z-2979-2020 Class II Terminated

Recalled by Life Technologies Corporation — Pleasanton, CA

FDA device recall Z-2979-2020 was initiated by Life Technologies Corporation on Aug 6, 2020 and is designated Class II. Reason for recall: COVID-19 Interpretive Software issues that may cause: 1) the risk of invalid or inconclusive tests/false negative and potentially increasing the retesting burden on customers. 2) false positive… The recall status is terminated (terminated May 5, 2022).

Recall details

Product Type
Devices
Report Date
Sep 23, 2020
Initiation Date
Aug 6, 2020
Termination Date
May 5, 2022
Voluntary/Mandated
Voluntary: Firm initiated

Product description

The TaqPath RT-PCR COVID-19 Kit with Applied Biosystem's COVID-19 Interpretive Software v2.0 or v2.2 (used with the QuantStudio Real-Time PCR platforms), includes the assays and controls for a multiplex real-time RT-PCR test for the qualitative detection of RNA from SARS-CoV-2 in upper respiratory specimens and bronchoalveolar lavage specimens from individuals suspected of COVID-19 by their healthcare provider. The test is interpreted by COVID-19 Interpretive Software

Reason for recall

COVID-19 Interpretive Software issues that may cause: 1) the risk of invalid or inconclusive tests/false negative and potentially increasing the retesting burden on customers. 2) false positive results due to improper vortexing.

Distribution pattern

US: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NM, NY, NV, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. OUS: Afghanistan, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Botswana, Brazil, Cameroon, Canada, Cypress, Estonia, France, Georgia, Germany, Greece, Guadeloupe, Hungary, India, Indonesia, Ireland, Israel, Italy, Kenya, Kuwait, Lebanon, Mozambique, Namibia, New Zealand, Oman, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, United Republic of Tanzania, and Vietnam.

Code information

Product SKU: Product Lot/Serial#: Release Date: 100093765 Version 1.2 (EUA) 28-APR-20 100087427 Version 2.2 (EUA) 11-MAY-20 100093771 Version 1.2 (CE-IVD) 28-APR-20 100094318 Version 2.2 (CE-IVD) 19-MAY-20 Applied Biosystems" COVID-19 Interpretive Software Version 1.2 was the first version of software developed in a dedicated CE-IVD Edition. Previous versions are shared with the FDA-EUA version of the device.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2979-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Life Technologies Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2979-2020. Report a data issue: hello@regdatalab.com.