Z-3005-2020 Class II Terminated
FDA device recall Z-3005-2020 was initiated by Siemens Medical Solutions USA, Inc on Aug 19, 2020 and is designated Class II. Reason for recall: Calibration offset between the central CT image plane and the external RT lasers has been set automatically to the factory default of 50.00 cm, if the misadjustment of the offset value is not… The recall status is terminated (terminated Apr 19, 2023). Affected quantity: 64 systems US.
Recall details
- Product Type
- Devices
- Report Date
- Sep 30, 2020
- Initiation Date
- Aug 19, 2020
- Termination Date
- Apr 19, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 64 systems US
Product description
Reason for recall
Distribution pattern
US Nationwide distribution.