Z-3012-2026 Class II Ongoing
FDA device recall Z-3012-2026 was initiated by Midmark Corporation on July 20, 2026 and is designated Class II. Reason for recall: A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation. The recall status is ongoing. Affected quantity: 3624 units.
Recall Details
- Product Type
- Devices
- Report Date
- September 2, 2026
- Initiation Date
- July 20, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3624 units
Product Description
Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.
Reason for Recall
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
Distribution Pattern
Worldwide distribution - US Nationwide and the country of Canada.
Code Information
Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark IQecg Digital ECG. The Midmark IQecg Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 480000 4820704 UDI-DI: 00841709100581, 00841709100321, 00841709100895, 00841709100901