Z-3045-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 19, 2018
- Initiation Date
- July 19, 2018
- Termination Date
- May 23, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 800 units
Product Description
Beaver Accu-Temp High Temperature Cautery, 2 inch Shaft with Fine Tip, Model Number 8445000 Product Usage: The indications and intended use of the High Temp cautery are for coagulating tissue or arresting bleeding from small vessels by heat conducted through the wire tip. Indications include ophthalmic, general and plastic surgery.
Reason for Recall
A small number of devices in this lot may have a melted cap and a hole formed through the packaging. The risk is potentially the product will no longer be sterile for use and may cause harm to a patient.
Distribution Pattern
Worldwide Distribution - US Nationwide in the states of: CA, DC, FL, IA, ID, IL, MA, MI, MN, NY, OK, PA, TN, TX, UT, and WA and the countries of Australia, France, Netherlands, Spain, and United Kingdom.
Code Information
Lot 6002809, UDI 8445000 30886158010297