Z-3045-2018 Class II Terminated

Recalled by Beaver Visitec — Waltham, MA

FDA device recall Z-3045-2018 was initiated by Beaver Visitec on Jul 19, 2018 and is designated Class II. Reason for recall: A small number of devices in this lot may have a melted cap and a hole formed through the packaging. The risk is potentially the product will no longer be sterile for use and may cause harm to a… The recall status is terminated (terminated May 23, 2019). Affected quantity: 800 units.

Recall details

Product Type
Devices
Report Date
Sep 19, 2018
Initiation Date
Jul 19, 2018
Termination Date
May 23, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
800 units

Product description

Beaver Accu-Temp High Temperature Cautery, 2 inch Shaft with Fine Tip, Model Number 8445000 Product Usage: The indications and intended use of the High Temp cautery are for coagulating tissue or arresting bleeding from small vessels by heat conducted through the wire tip. Indications include ophthalmic, general and plastic surgery.

Reason for recall

A small number of devices in this lot may have a melted cap and a hole formed through the packaging. The risk is potentially the product will no longer be sterile for use and may cause harm to a patient.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of: CA, DC, FL, IA, ID, IL, MA, MI, MN, NY, OK, PA, TN, TX, UT, and WA and the countries of Australia, France, Netherlands, Spain, and United Kingdom.

Code information

Lot 6002809, UDI 8445000 30886158010297

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3045-2018
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-3045-2018. Report a data issue: hello@regdatalab.com.