Z-3049-2017 Class II Terminated

Recalled by Alcon Research, Ltd. — Fort Worth, TX

FDA device recall Z-3049-2017 was initiated by Alcon Research, Ltd. on Jun 30, 2017 and is designated Class II. Reason for recall: Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens… The recall status is terminated (terminated Aug 27, 2019). Affected quantity: 429 units.

Recall details

Product Type
Devices
Report Date
Aug 30, 2017
Initiation Date
Jun 30, 2017
Termination Date
Aug 27, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
429 units

Product description

ORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgery

Reason for recall

Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.

Distribution pattern

Worldwide distribution. US Nationwide and Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, France, India, Japan, Mexico, Netherlands, Panama, Portugal, Romania, Spain, Thailand, UAE, and United Kingdom

Code information

Serial numbers: C5013,C5014,C5015,C5016,C5017,C5018,C5019,C5020,C5021,C5023,C5024,C5025,C5026,C5027,C5028,C5029,C5030,C5031,C5032,C5033,C5034,C5035,C5036,C5038,C5039,C5040,C5041,C5042,C5043,C5044, C5045,C5046,C5047,C5048,C5049,C5050,C5051,C5053,C5054,C5055,C5059,C5060,C5061,C5062,C5064, C5065,C5066,C5069,C5071,C5075,C5076,C5077,C5078,C5079,C5080,C5081,C5082,C5083,C5090,C5093, C5094,C5097,C5098,C5101,C5102,C5105,C5106,C5107,C5108,C5109,C5110,C5111,C5112,C5113,C5116, C5117,C5119,C5120,C5121,C5122,C5125,C5126,C5127,C5128,C5129,C5130,C5131,C5132,C5133,C5134, C5136,C5137,C5138,C5139,C5140,C5141,C5142,C5143,C5144,C5146,C5147,C5148,C5149,C5150,C5151, C5152,C5153,C5154,C5155,C5156,C5157,C5158,C5159,C5160,C5161,C5163,C5164,C5165,C5166,C5167, C5168,C5169,C5170,C5171,C5173,C5174,C5175,C5176,C5177,C5178,C5179,C5180,C5181,C5182,C5183, C5184,C5185,C5186,C5187,C5188,C5189,C5190,C5191,C5192,C5193,C5194,C5195,C5196,C5197,C5198, C5199,C5200,C5201,C5202,C5203,C5204,C5205,C5207,C5214,C5215,C5216,C5217,C5218,C5219,C5220, C5221,C5222,C5223,C5224,C5225,C5226,C5227,C5228,C5229,C5230,C5231,C5232,C5233,C5238,C5239, C5240,C5241,C5242,C5243,C5244,C5245,C5246,C5248,C5249,C5250,C5251,C5252,C5253,C5255,C5256, C5257,C5258,C5259,C5260,C5261,C5262,C5263,C5264,C5265,C5266,C5267,C5268,C5269,C5270,C5271, C5272,C5273,C5274,C5276,C5277,C5278,C5279,C5280,C5281,C5282,C5283,C5284,C5285,C5286,C5287, C5288,C5289,C5290,C5291,C5292,C5293,C5294,C5295,C5296,C5297,C5298,C5299,C5300,C5301,C5302, C5303,C5304,C5305,C5306,C5307,C5308,C5309,C5310,C5311,C5312,C5313,C5314,C5315,C5316,C5317, C5318,C5319,C5320,C5321,C5322,C5323,C5326,C5328,C5329,C5330,C5334,C5335,C5336,C5339,C5340, C5341,C5343,C5348,C5349,C5350,C5352,C5353,C5354,C5355,C5356,C5357,C5358,C5359,C5360,C5361, C5362,C5363,C5364,C5365,C5366,C5368,C5369,C5370,C5371,C5372,C5373,C5374,C5375,C5376,C5377, C5378,C5379,C5380,C5381,C5382,C5383,C5384,C5385,C5386,C5387,C5388,C5389,C5390,C5391,C5392, C5393,C5394,C5395,C5396,C5397,C5398,C5399,C5400,C5401,C5402,C5403,C5404,C5405,C5406,C5407, C5408,C5409,C5410,C5411,C5412,C5413,C5414,C5415,C5416,C5417,C5418,C5434,C5435,C5436,C5437, C5438,C5439,C5440,C5441,C5442,C5443,C5444,C5446,C5448,C5449,C5451,C5452,C5453,C5454,C5456, C5458,C5462,C5463,C5464,C5465,C5466,C5467,C5469,C5471,C5473,C5475,C5476,C5477,C5478,C5479, C5484,C5485,C5486,C5487,C5488,C5489,C5490,C5493,C5494,C5495,C5496,C5497,C5499,C5500,C5501, C5502,C5503,C5507,C5508,C5509,C5510,C5511,C5512,C5515,C5516,C5517,C5518,C5519,C5520,C5521, C5522,C5524,C5525,C5526,C5527,C5528,C5529,C5530,C5532,C5533,C5534,C5535,C5536,C5537,C5538, C5539,C5544,C5545,C5546,C5547,C5549,C5550,C5553,C5555

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3049-2017
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Alcon Research, Ltd.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-3049-2017. Report a data issue: hello@regdatalab.com.