Z-3059-2026 Class II Ongoing
Recalled by Medline Industries, LP
— Northfield, IL
FDA device recall Z-3059-2026 was initiated by Medline Industries, LP on Jun 12, 2026 and is designated Class II. Reason for recall: The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the… The recall status is ongoing. Affected quantity: 3752 kits.
Recall details
- Product Type
- Devices
- Report Date
- Sep 9, 2026
- Initiation Date
- Jun 12, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3752 kits
Product description
Medline Convenience Kits: 1) LABOR CDS, Model Number: CDS981762B; 2) VAG DELIVERY ADMISSION BAG, Model Number: DYKA1054F
Reason for recall
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
Distribution pattern
US Nationwide distribution.
Code information
1) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 25ABK513; 2) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 25DBM841; 3) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 25EBU475; 4) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 25FBQ817; 5) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 25HBK086; 6) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 23LBV280; 7) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 24BBN930; 8) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 24DBK663; 9) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 24DBS890; 10) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 24FBH870; 11) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 24GBR415; 12) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 24HBK593; 13) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 24JBE456; 14) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 24JBN932; 15) CDS981762B, UDI-DI: 10193489306767(each), 40193489306768(case), Lot Number: 23KBQ478; 16) DYKA1054F, UDI-DI: 10193489312607(each), 40193489312608(case), Lot Number: 25DLA351; 17) DYKA1054F, UDI-DI: 10193489312607(each), 40193489312608(case), Lot Number: 25ELA624; 18) DYKA1054F, UDI-DI: 10193489312607(each), 40193489312608(case), Lot Number: 24BLA086; 19) DYKA1054F, UDI-DI: 10193489312607(each), 40193489312608(case), Lot Number: 24CLA271; 20) DYKA1054F, UDI-DI: 10193489312607(each), 40193489312608(case), Lot Number: 24DLA466; 21) DYKA1054F, UDI-DI: 10193489312607(each), 40193489312608(case), Lot Number: 24FLA328; 22) DYKA1054F, UDI-DI: 10193489312607(each), 40193489312608(case), Lot Number: 24GLA526; 23) DYKA1054F, UDI-DI: 10193489312607(each), 40193489312608(case), Lot Number: 23JLB277; 24) DYKA1054F, UDI-DI: 10193489312607(each), 40193489312608(case), Lot Number: 23KLB020
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-3059-2026
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Medline Industries, LP? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works