Z-3062-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-3062-2026 was initiated by Medline Industries, LP on Jun 12, 2026 and is designated Class II. Reason for recall: The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the… The recall status is ongoing. Affected quantity: 18290 kits.

Recall details

Product Type
Devices
Report Date
Sep 9, 2026
Initiation Date
Jun 12, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18290 kits

Product description

Medline Convenience Kits: 1) ORTHO ASPIRATION KIT, Model Number: DYKM1636A; 2) SHOULD BLOCK KIT, Model Number: DYKM1881A; 3) TRIMA KIT, Model Number: DYNDB3077C

Reason for recall

The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.

Distribution pattern

US Nationwide distribution.

Code information

1) DYKM1636A, UDI-DI: 10889942925858(each), 40889942925859(case), Lot Number: 25ABW688; 2) DYKM1636A, UDI-DI: 10889942925858(each), 40889942925859(case), Lot Number: 23LBT438; 3) DYKM1881A, UDI-DI: 10198459025174(each), 40198459025175(case), Lot Number: 24KBS611; 4) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 25BBB106; 5) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24KBJ751; 6) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24LBO630; 7) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24ABB622; 8) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24CBG290; 9) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24CBG291; 10) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24CBT163; 11) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24EBM679; 12) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24FBL963; 13) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24HBV589; 14) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24HBV590; 15) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24HBV591; 16) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24HBV592; 17) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24HBV593; 18) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24JBV697; 19) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24JBV698; 20) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 24JBV699; 21) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 23IBT699; 22) DYNDB3077C, UDI-DI: 10193489273809(each), 40193489273800(case), Lot Number: 23JBS210

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3062-2026
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medline Industries, LP? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-3062-2026. Report a data issue: hello@regdatalab.com.