Z-3067-2017 Class II Terminated
FDA device recall Z-3067-2017 was initiated by Pentax Of America Inc on May 23, 2017 and is designated Class II. Reason for recall: Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the… The recall status is terminated (terminated May 21, 2020). Affected quantity: US: 118 units.
Recall details
- Product Type
- Devices
- Report Date
- Aug 30, 2017
- Initiation Date
- May 23, 2017
- Termination Date
- May 21, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- US: 118 units
Product description
Reason for recall
Distribution pattern
Nationwide
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-3067-2017
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for Pentax Of America Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Pentax Of America Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.