Z-3068-2017 Class II Terminated

Recalled by Pentax Of America Inc — Montvale, NJ

FDA device recall Z-3068-2017 was initiated by Pentax Of America Inc on May 23, 2017 and is designated Class II. Reason for recall: Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the… The recall status is terminated (terminated May 21, 2020). Affected quantity: US: 116 units.

Recall details

Product Type
Devices
Report Date
Aug 30, 2017
Initiation Date
May 23, 2017
Termination Date
May 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US: 116 units

Product description

Ultrasound Video Gastroscope, Model EG-3870UTK. The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.

Reason for recall

Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.

Distribution pattern

Nationwide

Code information

H110516, H110619, H110635, H110643, H110624, A110407, A110411, H110507, A110352, A110444, A110417, A110419, H110493, H110494, H110522, H110584, H110586, H110590, H110591, H110559, A110471, A110474, A110355, A110359, H110606, H110566, H110568, H110498, H110503, H110509, H110510, H110637, H110560, H110514, A110427, A110372, A110438, H110587, H110598, H110600, A110439, A110440, A110443, A110450, A110448, A110451, A110453, H110542, H110569, A110445, A110454, H110565, A110356, A110357, H110592, H110593, H110595, A110346, H110508, A110461, H110620, H110623, H110641, H110642, H110613, A110460, A110380, A110463, H110577, A110457, A110401, H110610, H110611, H110502, H110511, H110518, H110519, H110578, H110557, H110561, A110422, A110432, H110512, H110626, H110628, H110629, H110633, H110634, H110539, H110544, A110341, A110418, H110596, H110604, A110387, A110388, H110601, H110615, H110571, H110574, H110599, H110506, A110350, A110398, H110575, A110449, A110360, A110363, A110368, H110554, H110515, H110567, A110410, A110414, A110485, H110553, H110563, H110589, A110376, A110384, H110549, H110564, H110556, H110536, H110646, H110533, H110543, A110458, A110459, A110383, H110538, H110499, A110344, A110345, A110348, A110452, A110391, A110475, A110483, A110486, A110488, A110477, A110479, A110480, A110481, A110482, A110375, A110406, A110470, A110429, A110435, H110618, A110393, A110394, A110447, A110462, A110466, H110555, H110631, H110636, H110639, H110524, H110525, H110526, H110530, H110531, H110532, H110534, H110535, A110476, A110416, A110420, A110379, A110381, A110385, A110484, A110492, H110640, H110579, H110581, H110572, H110573, A110425, A110431, A110441, A110442, H110517, A110426, A110433, A110434, A110353, H110614, H110616, A110464, A110465, H110520, H110523, H110550.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3068-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Pentax Of America Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Pentax Of America Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Pentax Of America Inc →

Source: openFDA record for Z-3068-2017. Report a data issue: hello@regdatalab.com.