Z-3074-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

FDA device recall Z-3074-2026 was initiated by Medline Industries, LP on July 20, 2026 and is designated Class II. Reason for recall: Convenience kits contain BD ChloraPrep Clear - 1mL Applicators from specific lots that were recalled by the manufacturer because they may have an open or incomplete seal on the applicator packaging. U… The recall status is ongoing. Affected quantity: 222,653 total.

Recall Details

Product Type
Devices
Report Date
September 9, 2026
Initiation Date
July 20, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
222,653 total

Product Description

Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators DELUXE PRE OP KIT DYKS2018 GENERAL PREOP KIT DYKS1584A PRE OP KIT DYKS1216I PRE-OP KIT DYKM1398B

Reason for Recall

Convenience kits contain BD ChloraPrep Clear - 1mL Applicators from specific lots that were recalled by the manufacturer because they may have an open or incomplete seal on the applicator packaging. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.

Distribution Pattern

US Nationwide distribution.

Code Information

DYKS2018 UDI-DI 10198459188480 Lot 25GMI043 DYKS1584A UDI-DI 10198459146633 Lots 24JBA740 25ABF244 DYKS1216I UDI-DI 10198459362668 Lots 25GBP888 25JBD789 26ABL474 26CBF334 26DBD640 26EBB855 26FBC632 DYKM1398B UDI-DI 10198459413827 Lots 25HBI183 25JBK222 25KBA990 25LBJ990 26BBI525 26CBJ249 26EBQ630 26FMF189