Z-3155-2026 Class II Ongoing

Recalled by Philips North America Llc — Cambridge, MA

FDA device recall Z-3155-2026 was initiated by Philips North America Llc on Aug 18, 2026 and is designated Class II. Reason for recall: The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving… The recall status is ongoing. Affected quantity: 62.

Recall details

Product Type
Devices
Report Date
Oct 7, 2026
Initiation Date
Aug 18, 2026
Voluntary/Mandated
FDA Mandated
Product Quantity
62

Product description

Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Reason for recall

The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.

Distribution pattern

U.S. and U.S. Territories.

Code information

UDI: (01)00884838101111

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3155-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-3155-2026. Report a data issue: hello@regdatalab.com.