Z-3155-2026 Class II Ongoing
FDA device recall Z-3155-2026 was initiated by Philips North America Llc on Aug 18, 2026 and is designated Class II. Reason for recall: The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving… The recall status is ongoing. Affected quantity: 62.
Recall details
- Product Type
- Devices
- Report Date
- Oct 7, 2026
- Initiation Date
- Aug 18, 2026
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 62
Product description
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
Reason for recall
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
Distribution pattern
U.S. and U.S. Territories.
Code information
UDI: (01)00884838101111
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-3155-2026
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab