Z-3186-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 26, 2018
- Initiation Date
- April 30, 2018
- Termination Date
- April 17, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 272 KITS in total
Product Description
PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.
Reason for Recall
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
Distribution Pattern
US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.
Code Information
Lot : A735 Exp: 10/31/2018 UDI: (01)04987595301636(17)181031(10)A735 and Lot: O736 Exp: 01/03/2019 UDI: (01)04987595301636(17)190131(10)O736