Z-3209-2024 Class II Ongoing

Recalled by Microcare Medical — Commerce City, CO

Recall Details

Product Type
Devices
Report Date
October 2, 2024
Initiation Date
August 29, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1649

Product Description

Good Vibrations Ultrasonic Cleaner Solution, JorVet J0382G Ultrasonic Instrument Cleaning Solution, SpecClean Ultrasonic Cleaner Solution, Models: GVUS128-1, GVUS128-1EA, GVUS550-1, GVUS24-1, GVUS24-1EA, JGVUS128-1, VGVUSNP128-1, VGVUSNP128-1EA

Reason for Recall

Contaminated equipment and packaging caused Klebsiella oxytoca, Enterobacter asburiae, Enterobacter bugandensis, pathogenic and opportunistic bacteria, to occur in ultrasonic detergent, which caused product bloating/leaking, and if used during medical/dental procedures could result in urinary tract, respiratory, bloodstream infections; workers handling contaminated product/instruments also at risk

Distribution Pattern

Worldwide - US Nationwide distribution in the states of MN, GA, PA, WA, CO, NY, MD, CA, MI, NE, UT, OK, WI, NJ, IL, ME, CT, KS, TX, NC, TN, SC, MO and the countries of Canada.

Code Information

Lot/Model: 2331982/GVUS128-1EA, GVUS128-1, GVUS550-1; 2332204/GVUS128-1EA, JGVUS128-1, GVUS128-1, VGVUSNP128-1EA, VGVUSNP128-1; 2332393/GVUS128-1EA, VGVUSNP128-1, GVUS24-1EA, GVUS24-1, GVUS128-1 2432755/JGVUS128-1 2432989/GVUS24-1EA, VGVUSNP128-1EA, GVUS24-1, VGVUSNP128-1, JGVUS128-1 Model/UDI-DI: GVUS128-1/+H816GVUS128H, GVUS24-1/+H816GVUSU24%, GVUS550-1/+H816GVUS550G, VGVUS128-1/+H816VGVUSNP128A, JGVUS128-1/723754044532