Z-3209-2024 Class II Ongoing
FDA device recall Z-3209-2024 was initiated by Microcare Medical on August 29, 2024 and is designated Class II. Reason for recall: Contaminated equipment and packaging caused Klebsiella oxytoca, Enterobacter asburiae, Enterobacter bugandensis, pathogenic and opportunistic bacteria, to occur in ultrasonic detergent, which caused p… The recall status is ongoing. Affected quantity: 1649.
Recall Details
- Product Type
- Devices
- Report Date
- October 2, 2024
- Initiation Date
- August 29, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1649
Product Description
Good Vibrations Ultrasonic Cleaner Solution, JorVet J0382G Ultrasonic Instrument Cleaning Solution, SpecClean Ultrasonic Cleaner Solution, Models: GVUS128-1, GVUS128-1EA, GVUS550-1, GVUS24-1, GVUS24-1EA, JGVUS128-1, VGVUSNP128-1, VGVUSNP128-1EA
Reason for Recall
Contaminated equipment and packaging caused Klebsiella oxytoca, Enterobacter asburiae, Enterobacter bugandensis, pathogenic and opportunistic bacteria, to occur in ultrasonic detergent, which caused product bloating/leaking, and if used during medical/dental procedures could result in urinary tract, respiratory, bloodstream infections; workers handling contaminated product/instruments also at risk
Distribution Pattern
Worldwide - US Nationwide distribution in the states of MN, GA, PA, WA, CO, NY, MD, CA, MI, NE, UT, OK, WI, NJ, IL, ME, CT, KS, TX, NC, TN, SC, MO and the countries of Canada.
Code Information
Lot/Model: 2331982/GVUS128-1EA, GVUS128-1, GVUS550-1; 2332204/GVUS128-1EA, JGVUS128-1, GVUS128-1, VGVUSNP128-1EA, VGVUSNP128-1; 2332393/GVUS128-1EA, VGVUSNP128-1, GVUS24-1EA, GVUS24-1, GVUS128-1 2432755/JGVUS128-1 2432989/GVUS24-1EA, VGVUSNP128-1EA, GVUS24-1, VGVUSNP128-1, JGVUS128-1 Model/UDI-DI: GVUS128-1/+H816GVUS128H, GVUS24-1/+H816GVUSU24%, GVUS550-1/+H816GVUS550G, VGVUS128-1/+H816VGVUSNP128A, JGVUS128-1/723754044532