Z-3332-2026 Class II Ongoing

Recalled by Beckman Coulter Mishima K.K. — Sunto-Gun

FDA device recall Z-3332-2026 was initiated by Beckman Coulter Mishima K.K. on Aug 3, 2026 and is designated Class II. Reason for recall: Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument… The recall status is ongoing. Affected quantity: 492 units.

Recall details

Product Type
Devices
Report Date
Oct 7, 2026
Initiation Date
Aug 3, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
492 units

Product description

Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.

Reason for recall

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

Distribution pattern

Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.

Code information

Products containing affected software: Item Numbers / UDI-DI codes: Item Number: B96692 UDI-DI code: 14987666542965 Item Number: B96693 UDI-DI code: 14987666542972 Item Number: B12183 UDI-DI code: 14987666540671 Item Number: N3151400 UDI-DI code: Due to the age of the device and the fact that it has exceeded its storage period, the UDI number is not available.)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3332-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-3332-2026. Report a data issue: hello@regdatalab.com.