ZEVEX, INC. FDA Inspection Record

FEI 1000117172 · Inspections, Form 483s, warning letters and import refusals

ZEVEX, INC. is an FDA-registered facility (FEI 1000117172) located in Salt Lake City, UT, US. RegDataLab tracks 7 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Sep 17, 2018: Voluntary Action Indicated (VAI)
Inspections on record
7 (2 OAI, 3 VAI, 2 NAI)
Last OAI inspection
Jul 30, 2015
Warning letters
1 listed by FDA for this FEI, latest Apr 20, 2011
Recalls as the recalling firm
7, latest Feb 27, 2019
Import refusals
2

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 17, 2018Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Product rework procedures, Lack of or inadequate procedures (21 CFR 820.90(b)(2))
Apr 13, 2017No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Documentation (21 CFR 820.80(e))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 4 more
Jul 30, 2015Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
and 11 more
Aug 30, 2012Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Availability (21 CFR 820.180)
Lack of or inadequate DHR procedures (21 CFR 820.184)
Investigation of device failures (21 CFR 820.198(c))
and 2 more
Oct 14, 2011Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Documentation (21 CFR 820.100(b))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Dec 29, 2010Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
and 12 more
Jun 15, 2009No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 27, 2019Class IIZ-2053-2019MOOG Curlin Infusion Administration Set, REF 340-4130, Non-DEHP Tubing with Non-Vented Bag… and 1 more product
Administration sets leaked at the filter.
Nov 10, 2017Class IIZ-0697-2018MOOG Curlin Infusion Administration Set, Ref 340-4114, Non-DEHP Microbore Tubing with… and 5 more products
Particulate found on the Curlin spike.
Dec 23, 2015Class IIZ-0688-2016Moog Curlin Infusion, Ambulatory Infusion Pump The Curlin Ambulatory Infusion System…
Moog is recalling Curlin 4000, 6000, and PainSmart Infusion Pumps manufactured or serviced by MMDG between March 18, 2015 and November 6…
Nov 20, 2015Class IIZ-0415-2016Catalog No. INF0020-A  EnteraLite Infinity Enteral Pump Delivery Set with ENFit Connector…
Potential for leaking between the purple ENFit connector and the white transitional stepped connector due to cracking or breaking of the…
Nov 8, 2012Class IIZ-0294-2013EnteraLite Infinity Enteral Feeding Pump identified with the following list numbers…
MOOG Medical Devices Group is conducting a voluntary for a limited set of serial numbers of the EnteraLite Infinity and Infinity Orange…
Oct 12, 2012Class IIZ-0764-2013Infinity Enteralite Ambulatory Feeding Pumps, List number: INFKIT2 - Addendum to…
The firm is adding an addendum to the operator's manual for Infinity Enteralite Ambulatory Feeding Pumps pertaining to the Infinite Dose…
Sep 26, 2012Class IIZ-0072-2013***This recall is being conducted due to an incomplete recall for RES 57829*** Curlin…
This recall is the same issue for the FDA processed recall initiated by Moog Medical Devices on 1/25/11 which was related to customers…

FDA record score

4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 24, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

13 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAISep 17, 20186 x 0.251.5
Class II recall Z-2053-2019, this FEI as the recalling firmFeb 27, 20194 x 0.251
Class II recall Z-0697-2018, this FEI as the recalling firmNov 10, 20174 x 0.251
Class II recall Z-0688-2016, this FEI as the recalling firmDec 23, 201510 years or older, so it adds no points0
Class II recall Z-0415-2016, this FEI as the recalling firmNov 20, 201510 years or older, so it adds no points0
Inspection classified OAIJul 30, 201510 years or older, so it adds no points0
Class II recall Z-0294-2013, this FEI as the recalling firmNov 8, 201210 years or older, so it adds no points0
Class II recall Z-0764-2013, this FEI as the recalling firmOct 12, 201210 years or older, so it adds no points0
Class II recall Z-0072-2013, this FEI as the recalling firmSep 26, 201210 years or older, so it adds no points0
Inspection classified VAIAug 30, 201210 years or older, so it adds no points0
Inspection classified VAIOct 14, 201110 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsApr 20, 201110 years or older, so it adds no points0
Inspection classified OAIDec 29, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 24, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 24, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 24, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 24, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing ZEVEX, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check ZEVEX, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Zevex, Inc.
Location
Salt Lake City, UT (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.