FDA Inspection 938132 — Zevex, Inc.
The FDA completed an inspection of Zevex, Inc. on Jul 30, 2015 in Salt Lake City, UT. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 15 citations were issued during this inspection. Product area: Devices.
Inspection details
- Classification
- Official Action Indicated (OAI)
- End Date
- Jul 30, 2015
- Fiscal Year
- 2015
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Salt Lake City, UT (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 14716 | 21 CFR 820.30(f) | Design verification - output does not meet input requirement |
| 14720 | 21 CFR 820.50(a)(3) | Acceptable supplier records, inadequate records |
| 3121 | 21 CFR 820.80(b) | Lack of or inadequate receiving acceptance procedures |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3149 | 21 CFR 820.180 | Availability |
| 3160 | 21 CFR 820.184 | Lack of or inadequate DHR procedures |
| 3282 | 21 CFR 820.90(a) | Nonconforming product, Lack of or inadequate procedures |
| 3345 | 21 CFR 820.200(a) | Servicing - Lack of or inadequate procedures |
| 3432 | 21 CFR 820.75(b)(2) | Documentation of validated process performance |
| 3677 | 21 CFR 820.30(g) | Design validation - software validation not performed |
| 3678 | 21 CFR 820.30(g) | Design Validation - Risk analysis not performed/inadequate |
| 3696 | 21 CFR 820.100(b) | Documentation |
| 546 | 21 CFR 820.75(a) | Lack of or inadequate process validation |
| 732 | 21 CFR 803.50(a)(2) | Individual Report of Malfunction |